Model 4325B Intelligent Digital Signal Conditioner, TEDS-Tag Auto ID
CellMite® Model 4325B digital signal conditioner connects to a strain gage-based load cell, extensometer, or pressure sensor and simultaneously generates a serial digital output for connection to a PC as well as a +/- 10V analog output.
Contact | Link To Product | More Products By This Company
Images
Click on an image to see larger size image ...
Specifications
Excitation: Voltage | : | 5 VDC. Nom. Load 350 ohms |
Internal Resolution | : | 24-bit |
Input Range | : | +/-5.5 mV/V |
Conversion Rate | : | 60 per second |
Error | : | 0/01%, +/- 1 count |
Precision Shunt | : | 60K |
Analog Output | : | 16-bit, Scalable, +/-10V |
Serial Data | : | Multi-drop RS232 |
Dimensions | : | 5.5' x 2.75" x 1.2" |
Weight | : | 8.3 oz (235g) |
Power | : | 5VDC, 150mA. (Included Adapter) |
Voltage | : | 110/240VAC, 60/50Hz |
Details
CellMite® 4325B Intelligent Digital Signal Conditioner with TEDS-Tag® Auto Identification
CellMite® Model 4325B is a compact digital signal conditioner that connects directly to a standard strain gage-based load cell, extensometer, or pressure sensor and simultaneously generates a serial digital output for connection to a PC as well as a +/- 10V analog output.
CellMite® Model 4325B compensates for transducer nonlinearities using its integrated 6-point calibration capability not found in standard signal conditioners. Its multi-drop RS232 serial port, unit addressing capabilities, and simple command set allow for connection of multiple units in a serial network configuraton. The TEDS-Tag® auto identification feature with integrated electronic tags allows the CellMite® Model 4325B to automatically select the correct sensor calibration and analog output for scaling for the connected transducer.
CellMite® Model 4325B features a nonvolatile memory for parameter and calibration storage, the ability to select between three transducers, multi-point and mV/V calibration, remote sense excitation, and a 24-bit internal resolution with 16-bit analog output.
CellMite® Model 4325B Features:
Single Channel (Model 4325B)
2-pt mV/V Calibration
6-pt Linearization Calibration
Scalable +/-10V Analog Output
16-Bit Analog Output
Direct to PC with RS232 Data
24-Bit Resolution
Multiple Filter Selections
Storage for Three Calibrated Transducers
Remote Sense Excitation
Addressable for Serial Network
Tare, Peak, Valley
Unit Support: In, Cm, mm, %, Lb, Kg, N, mV/V, User Defined
TEDS-Tag® Auto Identification
Power Supplies Internally Isolated
Graphical User Interfaces and Software Drivers sold separately. Select from:
CellView Lite GUI (Cat. No. 514569)
CellView Multi-Display (Cat. No. 514424)
CellView with Graphing GUI (Cat. No. 514358)
Software Driver (Cat. No. 514864)
CellMite® Model 4325B Specifications:
CellMite Model 4325B Specifications:
Excitation: 5 VDC, Nom. Loac 350 ohms.
Operation:
Internal resolution: 24-bit
Input Range: +/-5.5 mV/V
Conversion rate: 60 per second.
Error: 0/01%, +/- 1 count.
Precision Shunt 60K.
Outputs:
Analog: 16-bit, Scalable, +/-10V.
Serial Data: Multi-drop RS-232.
Mechanical:
Size: 5.5" x 2.75" x 1.2"
Weight: 8.3 oz. (235g).
Power:
5 VDC, 150mA, (Included Adapter)
110/240VAC, 60/50 Hz.
Included Items;
DB9 Mating Connectors
Operational Manual on CD ROM.
Check out our complete line of CellMite® Intelligent Digital Signal Conditioners
CellView Lite GUI Software (Cat. No. 514569)
The CellView Lite Windows-based Software provides an easy graphical user interface (GUI) to CellMite® products.
CellView Lite Software Features:
Easy guides through adding/removing CellMite® units.
Stores calibration data for three sensors.
Guides user through sensor calibration.
Ability to individually name CellMite® units.
Tare and reset peak and valley for sensor.
Set output data with a quadratic filter.
Save/Load the CellMite® unit and sensor setup information.
Setup a test to start and stop automatically.
LEDs indicate operational status of the CellView Lite interface.
Save date, calibration, and sensor test information to spread sheets.
Control output with two relay switches and analog output of voltage.
Units of Lb, Kg, In, Cm, %, and User-defined.
CellView Multi-Display GUI Software (Cat. No. 514424)
The CellView Multi-Display Graphical User Interface (GUI) application communicates with up to eight precision CellMite® digitizing signal conditioners that are designed to amplify, filter and linearize the output of a strain gage sensor. CellView Multi-Display presents the user with a graphical user interface to monitor incoming load data from the CellMite® unit and to visualize and refine the data.
CellView Multi-Display GUI Software Features:
User interface connects up to eight pre-addressed CellMite® Strain Gage or CellMite® LVDT units.
Units can hold calibration data for up to three separate sensors.
User interface guides user through sensor calibrations.
User interface allows creation of up to eight independent displays of data from the various connected units.
Each display relates to a specific sensor.
User can tare and reset peak and valley for the sensor from the displays.
User can set filter level of the sensor via the unit, and/or further filter data as it enters the GUI software.
User may manually start and stop tests or automate the process.
Tests may be run in segments where each segment has a different data frequency.
User can save and load CellMite®, Sensor, and Test setup info.
User may save data and/or test info for use with spread sheets.
User can refine data as it comes into the Unit by applying a 5th order polynomial.